Welcome to our new Instron Community Blog hosted by Instron. It is a compilation of the freshest, brightest, most-talented minds that Instron has to offer. The world of materials science is so vast and encompasses the broadest range of industries, materials, and challenges that no one person can possibly possess all the knowledge required to be the resident expert – or master of materials science. It takes a small army behind the scenes collaborating and sharing technical know-how, experiences, and ideas to present the most accurate, relevant, and timely information to you – our readers.

We invite you to tell us who you are, share your stories and talk about your experiences. Join the Instron Community.

Thursday, December 13, 2012

Balancing the Load Cell

We are often asked how frequently an operator should balance the load cell during testing. Many lab managers require the operator to balance a load cell before the start of a new sample; others require balancing the load cell before testing every specimen. We believe that either procedure is acceptable, as long as you follow one major rule:

 
Never balance the load cell when there is a specimen clamped in both grips

Instron load cells can detect a change in load as a result of simply gripping the specimen. If the load is balanced after a specimen is gripped, you risk zeroing out a real load. This real load will be subtracted from or added to reported results, thereby falsely increasing or decreasing actual values depending on whether or not there was a compressive or tensile load on the specimen before the load cell was balanced.

If you notice a change on the load channel display after gripping, you can remove the load using automated software features such as preload or specimen protect, or by manually adjust the position of the crosshead.

3 comments:

Frank Lio said...

Another option to corect for speciemn tension or slack after gripping a specimen is to let the software do the correction. Most software should be able to extrapolate the true "0,0" (zero load at zero displacement) point via correction calculations. These calculations apply a correction to the specimen to compensate for any preload or slack in the
specimen and is always the first calculation and subsequent calculations use
the corrected gauge length. Very pertinent to ASTM D638's Toe Compensation (Annex A1).