Welcome to our new Instron Community Blog hosted by Instron. It is a compilation of the freshest, brightest, most-talented minds that Instron has to offer. The world of materials science is so vast and encompasses the broadest range of industries, materials, and challenges that no one person can possibly possess all the knowledge required to be the resident expert – or master of materials science. It takes a small army behind the scenes collaborating and sharing technical know-how, experiences, and ideas to present the most accurate, relevant, and timely information to you – our readers.

We invite you to tell us who you are, share your stories and talk about your experiences. Join the Instron Community.

Friday, September 13, 2013

What Does Load Measurement Accuracy of 0.5% of Reading Down to 1/100th, 1/500th, 1/1000th of a Load Cell's Capacity Imply?


First of all, there are two important pieces of information here:

(1) Load accuracy: ± 0.5% of the reading

(2) Lowest possible load within this accuracy: 1/100, 1/500, 1/1000 of load cell capacity

For example, let’s assume your test system has a load measurement accuracy of ± 0.5% of reading down to 1/100th of the load cell’s capacity. When a load cell of 10 000 N (10 kN) capacity is used on this test system and a measurement of 1000 N is read, the actual load measurement can be any value from 995 N to 1005 N. For this 10 kN capacity load cell, the minimum possible load that is guaranteed to meet the load accuracy of ± 0.5% of reading is 100 N.


Now, imagine you have a newer system with a new 10 kN load cell and the load measurement accuracy of ± 0.5% of reading is now extended down to 1/1000th of the load cell’s capacity. The load measurement accuracy of ± 0.5% of reading remains, but now the minimum possible load that is guaranteed to meet the load accuracy of ± 0.5% of reading is extended down to 10 N.

So what does a larger load accuracy range mean to you? Here are some of the benefits you will see:

- Less initial cost since less load cells are needed

- Lower subsequent service costs as fewer load cells need to be verified annually

- Less operator error and improved throughput (number of specimens tested per hour) by not having to change load cells between different test types such as tensile and flexure tests.

- Higher confidence in test data for low load tests.